5.9
Printing Results
The Report button allows you to print both on-screen graphs displays and current and voltage harmonics
test result data to an MS Word document. This provides a test report that displays the test setup, the Pass
or Fail result and a graph of the voltage and current waveform as well as the worst case values for each of
the current harmonics. The Report button is available at the end of a test run or any time a test is aborted
with the Stop button.
Information included on the harmonic test report will be a function of the selected test mode and EUT
connection stage. The following fields provide this information:
Heading:
Test category: Selected Stage and EUT type.
Current Test Result Summary
Test Result Overall Pass or Fail result
Highest parameters recorded during the test.
V_RMS EUT Voltage Line to Neutral
I_Peak Maximum peak current recorded
I_Fund Maximum fundamental current recorded (I1)
Power Maximum power level recorded
I_RMS Maximum RMS current recorded
Crest Factor
Power Factor
Harmonic Currents Table:
Harmonic number (#) Only harmonics for which limits apply are shown.
Harms (Avg) Average harmonics value (1.5 sec smoothing
filter applied)
100% Limit Absolute current value of 100 % limit for each
harmonic
% of Limit Percent of 100 % limit for each harmonic
Harms (Max) Maximum harmonic value recorded during test.
150% Limit Absolute current value of 150 % limit for each
harmonic
% of Limit Percent of 150 % limit for each harmonic
maximum
Status Individual harmonics pass or fail result (based on
100% and 150% limits)
Voltage Source Verification Data
The third page of the report contains data on the source voltage. If needed, the CTSMXL software will
compensate for voltage distortion that would affect the harmonic current results. This will be indicated.
Note that the overall EUT test results is repeated in the header of this page. This pass or fail does not
apply to the AC source.