Ametek ReFlex Programming Manual Bedienungsanleitung Seite 116

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 134
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 115
ReFlex Power™ Programming Manual Load Module Remote Programming
112 M380056-03 Rev L
R
EADBACK
F
UNCTIONS
Actual measured voltage, power and current
Voltage, current and resistance settings
Soft voltage, power and current limits
Status and Accumulated Status registers
Programming error codes
Fault codes
Manufacturer, power supply model, serial number, and firmware
version identification
5.3
POWER-ON CONDITIONS
Table 5-1 presents the power-on conditions for the Active Loads.
Table 5-1. Load Power-on Conditions
Condition Default
Current
0 Amps (initial power–on current). Also see
CAL<n>:INIT:CURR
Soft Voltage Limit Model maximum voltage
Soft Current Limit Model maximum current
OVP Trip Voltage
Model maximum voltage +7% (initial power–on OVP).
See
CAL<n>:INIT:VOLT:PROT
Input OFF, relays open
Grouped No Modules Grouped (example paralleled)
Soft Power Limit Model power limit
Seitenansicht 115
1 2 ... 111 112 113 114 115 116 117 118 119 120 121 ... 133 134

Kommentare zu diesen Handbüchern

Keine Kommentare