Ametek BPS Series Programming Manual Bedienungsanleitung Seite 44

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 226
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 43
AMETEK Programmable Power BPS / MX / RS Series SCPI Programming Manual
44 Manual P/N 7003-961 Rev AA
4.2.4 Voltage Measurement
CALibrate:MEASure:VOLTage[:AMBient][:AC][:FSCale] <NRf>
This command initiates the calibration of the rms voltage measurement at full scale and at
ambient temperature.
Command Syntax CALibrate:MEASure:VOLTage[:AMBient][:AC][:FSCale]
Parameters <NRf> (actual rms output voltage measured with external device)
Examples CAL:MEAS:VOLT 120
Query Syntax CALibrate:MEASure:VOLTage?
Returned Parameters <NR2> (value range -1000 to +1000)
Related Commands CALibrate:MEASure:VOLTage:TEMPerature:[:AC][:FSCale]
CALibrate:MEASure:VOLTage[:AMBient]:DC[:FSCale] <NRf>
This command initiates the calibration of the DC voltage measurement at full scale and at
ambient temperature.
Command Syntax CALibrate:MEASure:VOLTage[:AMBient]:DC[:FSCale]
Parameters <NRf> (actual DC output voltage measured with external device)
Examples CAL:MEAS:VOLT:DC 120
Query Syntax CALibrate:MEASure:VOLTage:DC?
Returned Parameters <NR2> (value range -1000 to +1000)
Related Commands CALibrate:MEASure:VOLTage:TEMPerature:DC[:FSCale]
CALibrate:MEASure:VOLTage[:AMBient]:DC:ZERO <NRf>
This command initiates the offset adjustment of the DC voltage measurement at ambient
temperature.
Command Syntax CALibrate:MEASure:VOLTage[:AMBient]:DC:ZERO
Parameters <NRf> (0 or desired offset value)
Examples CAL:MEAS:VOLT:DC:ZERO 0
Query Syntax CALibrate:MEASure:VOLT:DC:ZERO?
Returned Parameters <NR1> (value range -127 to +128)
Related Commands CALibrate:MEASure:VOLTage:TEMPerature:DC:ZERO
CALibrate:MEASure:VOLTage:TEMPerature[:AC][:FSCale] <NRf>
This command initiates the calibration of the rms voltage measurement at full scale and at
elevated temperature.
Command Syntax CALibrate:MEASure:VOLTage:TEMPerature[:AC][:FSCale]
Parameters <NRf> (actual rms output voltage measured with external device)
Examples CAL:MEAS:VOLT:TEMP 120
Query Syntax CALibrate:MEASure:VOLTage:TEMPerature?
Returned Parameters <NR2> (value range -1000 to +1000)
Related Commands CALibrate:MEASure:VOLTage[:AMBient][:AC][:FSCale]
Seitenansicht 43
1 2 ... 39 40 41 42 43 44 45 46 47 48 49 ... 225 226

Kommentare zu diesen Handbüchern

Keine Kommentare