Ametek PLA-PLW Programming Manual Bedienungsanleitung Seite 20

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 74
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 19
18
Returned Parameters <NRf+>: “CURR?” and “CURR:TRIG?” return the presently programmed
current levels. After a trigger signal or “ABORt”, “CURR:TRIG?” returns
the same value as “CURR?”.
“CURR? MAX”, “CURR? MIN”, “CURR:TRIG? MA X” and
“CURR:TRIG?” MIN return the maximum and minimum programmable
LEVel values.
[SOURce:]CURRent:[TRANsient:]DUTY
This command sets the duty cycle of each of the transients when the generator is in CONTinuous
mode.
Command Syntax CURRent:[TRANient:]DUTY <NR1>
Parameters <NR1>: Duty cycle value. 1 – 100.
Unit Percentage (%)
Query Syntax CURRent:DUTY?
CURRent:DUTY? MIN CURRent:DUTY? MAX
CURRent:TRANsient:DUTY?
CURRent:TRANsient:DUTY? MIN
CURRent:TRANsient:DUTY? MAX
Returned Parameters <NR1>, duty cycle value in percentage
[SOURce:]CURRent:[TRANsient:]FREQuency
This command sets the frequency of the transients when the generator is in CONTinuous mode.
Command Syntax CURRent:[TRANsient:]FREQuency <NRf+>
Parameters <NRf+>: Frequency value, refer to model specifications for range.
Unit Hertz
Query Syntax CURRent:FREQuency?
CURRent:FREQuency? MIN CURRent:FREQuency? MAX
CURRent:TRANsient:FREQuency?
CURRent:TRANsient:FREQuency? MIN
CURRent:TRANsient:FREQuency? MAX
Returned Parameters <NRf+>
[SOURce:]CURRent:PROTection[:LEVel]
[SOURce:]CURRent:PROTection:DELay
[SOURce:]CURRent:PROTection:OVER[:LEVel]
[SOURce:]CURRent:PROTection:OVER:DELay
[SOURce:]CURRent:PROTection:STATe
[SOURce:]CURRent:PROTection:UNDer[:LEVel]
[SOURce:]CURRent:PROTection:UNDer:DELay
[SOURce:]CURRent:PROTection:UNDer:STATe
Channel-Specific Current Limiting Command/Query. This command sets the over-current
protection limit or the under-current protection limit to the input current that the electronic load will sink.
When the input current reaches the over-current protection limit or falls below the under-current
protection limit for the specified delay period, the input of the electronic load is shut off and draws no
current.
The INPut:PROTection:CLEar command (or front panel key) re-enables the input current. The trigger
activated current functions (CURR[:LEV]:TRIG and CURR:TLEV) automatically keep track of
Seitenansicht 19
1 2 ... 15 16 17 18 19 20 21 22 23 24 25 ... 73 74

Kommentare zu diesen Handbüchern

Keine Kommentare