Ametek PLA-PLW Programming Manual Bedienungsanleitung Seite 31

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 74
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 30
29
MODE COMMANDS
[SOURce:]MODe
The commands sets and query the input regulation mode of the electronic load.
CURRent constant current mode
POWer constant power mode
RESistance constant resistance mode
VOLTage constant voltage mode
Command Syntax MODe <CRD>
Parameters CURR | POW | RES | VOLT
*RST Value CURR
Query Syntax MODe?
Returned Parameters <CRD> CURR= constant current mode
POW= constant power mode
RES= constant resistance mode
VOLT= constant voltage mode
[SOURce:]MODe:RANGe
The commands sets and query the operating range at the current operating mode.
Note: When Range control (“UTIL:RANG”) is set to AUTO (1), this command will cause
execution error.
Command Syntax MODe:RANGe <NR1>
Parameters range index 0 – Low Range
1 – Middle Range
2 – High Range
3 – Ultra High/Ultra Low Range
Query Syntax MODe:RANGe?
Returned Parameters <NR1>, range index number
Seitenansicht 30
1 2 ... 26 27 28 29 30 31 32 33 34 35 36 ... 73 74

Kommentare zu diesen Handbüchern

Keine Kommentare