Returned Parameters <NRf+> POW? and POW:TRIG? return the presently programmed
power levels. After a trigger or ABORt, POWer:TRIG? returns the same
value as POWer? .
POWer? MAX, POWer? MIN, POWer:TRIG? MAX and POWer:TRIG?
MIN return the maximum and minimum programmable LEVel and
TLEVel values.
[SOURce:]POWer:[TRANsient:]DUTY
This command sets the duty cycle of each of the transients when the generator is in CONTinuous
mode.
Command Syntax POWer:[TRANsient:]DUTY <NR1>
Parameters Duty cycle value. 1 – 100
Unit Percentage (%)
Query Syntax POWer: DUTY?
POWer: DUTY? MIN POWer: DUTY? MAX
POWer:TRANsient:DUTY?
POWer:TRANsient:DUTY? MIN POWer:TRANsient:DUTY? MAX
Returned Parameters <NR1> Duty cycle value in percentage.
[SOURce:]POWer:[TRANsient:]FREQuency
This command sets the frequency of the transients when the generator is in CONTinuous mode.
Command Syntax POWer:[TRANsient:]FREQuency <NRf+>
Parameters Frequency value, refer to model specifications for range | MIN | MAX
Unit Hertz
*RST Value 1.0 Hz
Query Syntax POWer:FREQuency?
POWer:FREQuency? MIN POWer:FREQuency? MAX
POWer:TRANsient:FREQuency?
POWer:TRANsient:FREQuency? MIN
POWer:TRANsient:FREQuency? MAX
Returned Parameters <NRf+>, frequency value.
[SOURce:]POWer:PROTection[:LEVel]
[SOURce:]POWer:PROTection:OVER[:LEVel]
[SOURce:]POWer:PROTection:DELay
[SOURce:]POWer:PROTection:OVER:DELay
[SOURce:]POWer:PROTection:UNDer:DELay
[SOURce:]POWer:PROTection:STATe
[SOURce:]POWer:PROTection:UNDer[:LEVel]
[SOURce:]POWer:PROTection:UNDer:STATe
Channel-Specific Power Limiting Command/Query. This command sets the over-power
protection limit or the under-power protection limit to the input power that the electronic load will sink.
When the input power reaches the over-power protection limit or falls below the under-power
protection limit for the specified delay period, the input of the electronic load is shut off and draws no
power.
The INPut:PROTection:CLEar command (or front panel key) re-enables the input power. The trigger
activated power functions (POWer[:LEV]:TRIG and POWer:TLEV) automatically keep track of