7.1.3 Lx/Ls Series ABD Option Limitations
The extent of coverage of the ABD option as implemented on the Lx/Ls Series platform is shown in
the table below. Tests marked with an „x‟ generally require additional hardware to be used. Tests
marked with a „P‟ require arbitrary waveform capability, which is available on Lx models or Ls models
with the ADV option only.
Note that the Lx/Ls Series only provides AC output so none of the DC tests (Table D and E) are
supported on the Lx/Ls Series.
Harmonic Measurement Restrictions
The maximum bandwidth for current harmonic measurements of the Lx/Ls Series is limited to 48KHz
in single-phase mode and 16 KHz in three-phase range. Harmonic components that fall outside this
bandwidth are not reported (value of 0.00 is returned). The same applies to voltage harmonics.
The requirement for ABD is 150 KHz so for EUT‟s with high harmonic order currents, a higher
bandwidth external power analyzer should be used in addition to the AC source.
No.
Description
Table
A
Table
B
Table
C
No.
Description
Table
D
No.
Description
Table
E
1
Steady State
Voltage
1
Steady State
Voltage
X
1
Steady State
Voltage
X
2
Abnormal Steady
State Voltage
2
Abnormal Steady
State Voltage
X
2
Voltage Surge -
Normal
X
3
Voltage Surge -
Normal
3
Voltage Surge -
Normal
X
3
Voltage Surge -
Abnormal
X
4
Voltage Surge -
Abnormal
4
Voltage Surge -
Abnormal
X
4
Voltage Spikes
X
5
Voltage Spikes
X
X
X
5
Voltage Spikes
X
5
Switching
Transients
Appendix 1
X
6
Switching
Transients
Appendix 1
6
Switching
Transients
Appendix 1
X
6
Ripple Voltage
X
7
Voltage Modulation
7
Square Wave
X
7
Starting Current
X
8
Normal Frequency
Variations
8
Ripple Voltage
X
9
Frequency
Modulation
10
Distorted Voltage
P
P
11
Voltage DC
Content
X
X
X
Table 7-3: -ABD Option Ls/Lx Test Coverage