User Manual - Rev H California Instruments
Avionics Software 195
8.4.1 Rev C:
8.4.1.1 Table SVF:
Steady State Voltage and Frequency
DO160 16.5.1.1.b, 620Hz and
390Hz
Surge to 160Vac max, Sag to 71Vac min
DO160 16.5.1.5.1.b, 620Hz and
390Hz
Vamp = 0.18V to 1.24V, Mrate = 1Hz to 200Hz
DO160 16.5.1.2, 620Hz and 390Hz
Requires Additional Equipment
Limits defined by Table A.4
Voltage Distortion - Endurance test with motor
Voltage Distortion - Endurance test without motor
Test requires multiple load cables
Test not performed with software
Slew from 390 to 620Hz, 100Hz/s up and down
Famp = 4.9 Hz to 0.2Hz, Mrate = 0.014Hz to
100Hz
DO160 16.5.1.3, 620Hz and 390Hz
Voltage Modulation due to Equipment
Test requires special cables
Test not performed with software
Voltage Spike Due to Equipment Load Switching
Requires Additional Equipment
Abnormal Steady State Voltage and Frequency
DO160 16.5.2.1.b, 620Hz and
390Hz
Abnormal Voltage Transients
Surge to 180Vac max, Sag to 0Vac min
DO160 16.5.2.2.b, 620Hz and
390Hz
Abnormal Voltage Modulation
Vamp = 0.18V to 1.24V, Mrate = 1Hz to 200Hz
DO160 16.5.1.2, 620Hz and 390Hz
Emergency Steady State Voltage and Frequency
DO160 16.5.1.1.b, 620Hz and
390Hz
Emergency Voltage Distortion
Voltage Distortion - Endurance test with motor
Voltage Distortion - Endurance test without motor
Emergency Voltage Distortion Transients
Emergency Frequency Variations
Slew from 390 to 620Hz, 500Hz/s up and down
Emergency Voltage Modulation due to Equipment
Test requires special cables
Test not performed with software